LACTOSCAN SCC
PN: SCCS LACTOSCAN SCC - PN: SCCS

Reference testing procedure save_alt Download File

To receive results with reference accuracy from somatic cell counting using LACTOSCAN SCC the following steps must be done:
Samples taking: Samples must be taken with accordance to the recommended sampling method given in ISO 707|IDF 50.
Samples storage: Samples must be stored with accordance to the recommendations in ISO 13366-1 | IDF 148-1:2008, page 4: “7.1 Storage Prior to testing or preservation, store the test samples at a temperature of 4 °C ± 2 °C.
Analyse the test samples within 6 h after sampling. In the case of longer storage, add chemical preservatives such as boric acid, bronopol or potassium dichromate. The final concentration of boric acid shall not exceed 0,6 g per 100 ml of test sample. The final concentration of bronopol shall not exceed 0,05 g per 100 ml of test sample. The final concentration of potassium dichromate shall not exceed 0,1 g per 100 ml of test sample. Store the thus preserved test samples at a temperature of 4 °C ± 2 °C for no longer than 6 days. For environmental reasons, it is recommended to restrict the use of potassium dichromate to samples that require a long shelf life only.” Note! Bronopol is the preferred preservative for preparing samples for analysis with LACTOSCAN SCC.


Other downloads for this product:

Brochure 2018-06-18 | 2.74MB save_alt
Folleto 2018-06-25 | 2.50MB save_alt
Брошюра 2018-06-25 | 1.85MB save_alt
Quick start guide 2018-06-25 | 5.82MB save_alt
User manual 2018-09-20 | 13.64MB save_alt
Manual de usuario 2018-08-28 | 13.66MB save_alt
Инструкция пользователя 2018-08-28 | 14.24MB save_alt
Инструкция за употреба 2018-09-20 | 13.36MB save_alt
Reference testing procedure 2018-06-25 | 502.28KB save_alt
Excel table for calculation of Coefficient of variation (CV %) 2018-07-26 | 11.49KB save_alt
Poster 2018-06-25 | 2.24MB save_alt
Póster 2018-06-25 | 2.26MB save_alt
Plakāts 2018-06-25 | 2.33MB save_alt
Плакат 2018-06-25 | 2.30MB save_alt
Working principle of SCC 2018-07-18 | 614.43KB save_alt
Comparison between Lactoscan Automated microscopic SCC and DMSCC 2019-01-14 | 4.98MB save_alt

MILKOTRONIC LTD
4, Narodni Buditeli Street
8900 Nova Zagora
BULGARIA
tel/fax: + 359 457 670 82
office@autocellcount.com

Copyright © 1997-2020 Autocellcount.com Milkotronic Ltd. All rights reserved. W