Excel table for calculation of Coefficient of variation (CV %) save_alt Download File

Coefficient of variation (CV) %:

Coefficient of variation (CV) % is measured in Reference mode following the instruction for Reference testing procedure. It is determined by analysis of standard reference milk samples from certified dairy laboratories. To determine the reference coefficient of variation (CV) % only such standard reference milk sample must be used. If nonstandard milk samples are used, a deviation of the CV% may be received. The standard reference milk samples do not include a large amount of cell fragments and do not include yeast cells.


Calculating Coefficient of variation (CV) %:

To calculate the Coefficient of variation (CV) %, following the above procedure, you have to test:

  • 10 times milk sample with somatic cell concentration 50000-200000 cell/ mL;
  • 10 times milk sample with somatic cell concentration 400000-600000 cell/mL;
  • 10 times milk sample with somatic cell concentration 800000-1400000 cell/mL;

Then fill in the results in the table below.


Other downloads for this product:

Brochure 2018-06-18 | 2.74MB save_alt
Folleto 2018-06-25 | 2.50MB save_alt
Брошюра 2018-06-25 | 1.85MB save_alt
Quick start guide 2018-06-25 | 5.82MB save_alt
User manual 2022-12-02 | 5.70MB save_alt
Manual de usuario 2018-08-28 | 13.66MB save_alt
Инструкция пользователя 2018-08-28 | 14.24MB save_alt
Инструкция за употреба 2018-09-20 | 13.36MB save_alt
Reference testing procedure 2018-06-25 | 502.28KB save_alt
Excel table for calculation of Coefficient of variation (CV %) 2018-07-26 | 11.49KB save_alt
Poster 2018-06-25 | 2.24MB save_alt
Póster 2018-06-25 | 2.26MB save_alt
Plakāts 2018-06-25 | 2.33MB save_alt
Плакат 2018-06-25 | 2.30MB save_alt
Working principle of SCC 2018-07-18 | 614.43KB save_alt
Comparison between Lactoscan Automated microscopic SCC and DMSCC 2019-01-14 | 4.98MB save_alt

MILKOTRONIC LTD
4, Narodni Buditeli Street
8900 Nova Zagora
BULGARIA
tel/fax: + 359 457 670 82
office@autocellcount.com

Copyright © 1997-2020 Autocellcount.com Milkotronic Ltd. All rights reserved. W